Quality & Traceability
The steel, the heat treatment, the masters and the decision rules behind every SVNM gauge
A gauge is only as good as the steel it is made from, the treatment that stabilises it and the master it is checked against. Because we manufacture gauges and run a NABL-accredited calibration laboratory (ISO/IEC 17025, certificate CC-2472) under one roof, we control all three.
Every gauge checked against a tighter standard
We work to two rules for every gauge we make:
- Rule 1: maximum deviation from nominal plus the uncertainty of the reference master must be within the specified gauge tolerance.
- Rule 2: the reference master must be accurate to at least 1/10th of the gauge tolerance.
| Gauge type | Reference master we use | Accuracy |
|---|---|---|
| Thread plug gauge (parallel) | Floating carriage diameter measuring machine (FCDMM) | 0.5 µm |
| Thread ring gauge (parallel) | Length measuring machine (LMM) | 0.1 µm |
| Air plug gauge | Master ring gauge, measured on LMM | 0.1 µm |
| Air ring gauge | Master plug gauge, measured on LMM | 0.1 µm |
| Master setting ring gauge | LMM | 0.1 µm |
| Plain plug gauge | LMM or comparator stand (by accuracy needed) | 0.1–2 µm |
| Plain ring gauge | LMM | 0.1 µm |
| Snap gauge | Slip gauges | 0.2 µm |
| Custom / special gauges | Coordinate measuring machine (CMM) | 1.0 µm |
See the full list of machines and reference standards on our infrastructure page.
A documented decision rule on every certificate
We do not just report a value. Before a certificate states that a gauge conforms, we apply the decision rule:
Maximum error + uncertainty of calibration ≤ acceptance criteria
This is why low uncertainty matters. Suppose a gauge is measured with an error of 6 µm:
Uncertainty ± 4 µm
The true error could be anywhere between 2 µm and 10 µm: a wide margin you cannot rely on.
Uncertainty ± 0.5 µm
The true error lies between 5.5 µm and 6.5 µm: a tight, dependable result.
Our decision rules follow:
- ISO 14253-1:2017 — decision rules for proving conformity or non-conformity with specifications
- ILAC-G8:2019 — guidelines on decision rules and statements of conformity
- NABL 141 — guidelines for estimating and expressing measurement uncertainty
The steel: OHNS, not EN-31
All our thread, air, master and plain gauges are made from OHNS (oil-hardening non-shrinking) tool steel from MUSCO (now SANYO), with a mill test report for every batch. Compared with EN-31 bearing steel, OHNS contains tungsten, which forms a hard carbide network that resists wear, and it moves very little when hardened.
| Element | EN-31 | OHNS (our choice) |
|---|---|---|
| Carbon (C) | 0.90–1.20% | 0.85–1.00% |
| Chromium (Cr) | 1.20–1.60% | 0.40–0.60% |
| Manganese (Mn) | 0.25–0.45% | 1.00–1.30% |
| Tungsten (W) | — | 0.40–0.60% |
The result is a gauge that wears more slowly, holds its size between calibrations and needs fewer re-calibrations over its life.
Heat treatment and sub-zero stabilisation
- Hardening to 60–62 HRC.
- Tempering to relieve stresses.
- Sub-zero treatment at −80 °C in our in-house freezer, to make the structure more uniform and stable.
- Tempering again in a controlled oven (up to 180 °C) to relieve the remaining stresses.
Sub-zero treatment converts retained austenite, so the gauge is less likely to change size over time and resists wear for longer. After grinding and lapping, every gauge is calibrated in our NABL-accredited laboratory and shipped with its certificate.
Frequently asked questions
Why do you use OHNS instead of EN-31 for gauges?
OHNS (oil-hardening non-shrinking tool steel, AISI O1 type) contains tungsten, which forms hard carbides that resist wear, and it moves very little on hardening. EN-31 is a bearing steel without tungsten. For a gauge that must hold its size for years, OHNS gives longer life and better dimensional stability.
What is a decision rule on a calibration certificate?
A decision rule states how measurement uncertainty is taken into account before a gauge is declared within or outside tolerance. We follow ISO 14253-1 and ILAC-G8: a gauge passes only when its maximum error plus the uncertainty of calibration stays within the acceptance limit.
How accurate are the masters used to check your gauges?
Our rule is that the reference master must be at least ten times more accurate than the gauge tolerance it checks. The table on this page shows the master and its accuracy for each gauge type.