How to handle, use and store gauges

Simple habits that keep your gauges accurate and your inspection results right

A gauge is only as reliable as the way it is used. These are the practices we follow in our own NABL-accredited laboratory and recommend to every customer.

1. Measure at 20 °C

  • Gauges and parts are specified at the international reference temperature of 20 °C (ISO 1). At any other temperature both grow or shrink.
  • Steel grows by about 11.5 µm per metre for every 1 °C. A 50 mm steel part at 25 °C is about 3 µm larger than at 20 °C, which can be a large share of a tight tolerance.
  • Keep the gauge and the part together on the inspection bench until both reach the same temperature. A part straight from the machine, or a gauge held in a warm hand, gives wrong results. Heavy parts need longer to settle.

2. Clean before every check

  • Wipe the gauge and the feature with a clean, lint-free cloth. Chips, coolant, oil films and burrs cause false rejections and scratch the gauge.
  • Remove burrs from the part before gauging; a burr can stop a good GO gauge or damage the gauge surface.

3. Apply the gauge correctly

  • Hold the gauge by its handle, not by the measuring surface, so body heat does not reach it.
  • Apply the gauge by hand under its own weight, without force. If you have to push a GO plug into a hole, the hole is undersize. Never use a hammer, wrench or extension on a gauge.
  • Check the full length of the feature with the GO gauge, and several positions with the NOGO gauge. NOGO ends are identified by a red marking.
  • Never gauge a part that is still moving, rotating or hot on the machine.

4. Store gauges safely

  • Return every gauge to its case after use. Do not let gauges touch each other or other tools.
  • Apply a thin film of rust preventive before storage, and store in a dry place away from magnets and coolant.
  • If a gauge is dropped, has a burr or shows rust or scratches on the gauging surface, take it out of use and have it checked before using it again.

5. Recalibrate on time

  • GO gauges wear with every check. Calibration shows whether a gauge still represents your drawing limits and how close it is to its wear limit.
  • Typical intervals are 3 to 6 months for gauges in continuous use and 6 to 12 months for normal use. Keep a record of each gauge, its certificate and its due date.
  • Use an ISO/IEC 17025 (NABL) accredited laboratory, so your certificates are accepted in customer and certification audits.

Frequently asked questions

Why must gauges be used at 20 °C?

20 °C is the international reference temperature for dimensional measurement (ISO 1). Gauges and parts are specified at 20 °C; at other temperatures they expand or contract, so measuring a warm part or with a warm gauge gives the wrong answer.

How should a GO/NOGO plug gauge be applied?

Clean the gauge and the hole, then apply the gauge by hand under its own weight without force. The GO end must pass through the full length of the hole and the NOGO end must not enter.

How should gauges be stored?

In their own cases, lightly coated with rust preventive, in a dry place, not touching other gauges or tools, and away from magnets and coolant.

What should I do if a gauge is dropped?

Take it out of use and have it checked or recalibrated before using it again, because a drop can damage the gauging surface or change its size.

Work out the size change for your part with our thermal expansion calculator, read what a GO/NOGO gauge is, or book a calibration.